Performance . The mechanical refrigerator blood bank shall be completely assembled in its operating configuration and tested to determine compliance with this specification. The refrigerator shall be operated to determine, proper functioning. The refrigerator shall not be defective. Testing shall include the following:
Operation/Inspection:
1. Interior chamber drain
2. Casters.
3. Moisture tight.
4. Seal and subsequent loss of cold air.
5. Door AJAR (The red LED flashes whenever the door is open or not completely closed).
6. Fail - Safe relay and external alarm adapter plug with “over/under-temp” alarm signal light.
7. Indicators and alarms.
8. Controls and switches.
9. Charts and ink.
10. Insulation
11. Battery test switch.
12. Power failure indicator light.
13. Cabinet: Inspect the cabinet arid parts thoroughly for any damage and missing parts. Inspect the interior/exterior of the refrigerator, including the mechanical equipment for special instructions and tags fastened at various points.
14. Line power fuse.
15. Motor overload protection.
16. Overall dimensions W X D X H